08069976663 /08069976665 chennaichemicals.com @ gmail.com

New In Store

Duis autem vel eum iriure dolor in hendrerit in vulputate velit esse molestie consequat, vel illum dolore eu feugiat
Shop Now

Latest News

Type what you are searching for:

Make your own custom-made popup window!

Lorem ipsum dolor sit amet, consectetuer adipiscing elit, sed diam nonummy nibh euismod tincidunt ut laoreet dolore
[contact-form-7 404 "Not Found"]

Product

Placeholder

Product for  

Silver Nitrate Soln. LR

This RXSOL standard solution make Colloidal Turbidity by refraction of the liquid. And Colloidal matter indicates hydrocarbons, chlorides and other residues.

Product Enquiry

Request for Quote and Get Heavy Discount on Price

Description

Laboratory GRADE materials ::: 

Chloride test:

50cc un-filtered wash add to 30 cc distilled water shake well to dissolve salt and then filter qualitatively.

Colloidal particles are not simple molecules but physical aggregation of molecules . They readily pass through ordinary filter paper.. Turbidity is a measure of the refraction of the liquid. Colloidal matter indicates hydrocarbons, chlorides and other residues.



Note: Methanol cannot dissolve salt or loosen the sodium cation/chloride anion from the NaCL ionic bond. -only water with high solvent power and high dielectric constant can do this. NaCL crystal is extremely strong and rigid. A Sodium ion is equally attracted to 6 Chloride ions.

To this 80cc mixture— add 1/2 cc silver nitrate( strength 10%) plus 2cc nitric acid ( strength 20%) plus 17.5 cc di water to make 100 cc total.

Shake and stand for 5 minutes. Compare milkiness with a standard comparator of 1ppm made as follows:

50cc Methanol + 4cc standard Chloride soln ( strength 10ppm ) + 1/2 cc Silver Nitrate ( strength 10%) + 2cc Nitric acid ( strength 20%) + 43.5 cc DI water to make total 100cc solution . Memorise this milkiness as the failing standard for the current voyage.

(for 0.5 ppm comparator put 2 cc of standard chloride soln, for 0.250 ppm put 1 CC.) For 2 ppm put 8cc NaCl —-for 4ppm put 16cc NaCl.



Why Nitric acid ??— Nitric acid parts the silver vide the assaying process as chemically silver is not active and dissolves only in strong Nitric acid or H2So4.

The method of not involving Nitric acid in the Chloride test could pass a ship staff tank WWT , while a shore gas ion chromatograph will fail the sample for Mono Ethylene Glycol fibre grade. Nitric acid decomposes Sodium Cyanide and Sodium Sulphide which otherwise form precipitate with AgNO3 and thus interfere with the lab test.

Show Buttons
Hide Buttons